Comparison of the Effects of Different Fixation Methods on Fragments and Temporomandibular Joint in Sagittal Split Ramus Osteotomy Applied to Patients with Mandibular Asymmetry Using Three-Dimensional Finite Element Analysis
PDF
Cite
Share
Request
Research Article
VOLUME: 10 ISSUE: 1
P: 44 - 50
February 2025

Comparison of the Effects of Different Fixation Methods on Fragments and Temporomandibular Joint in Sagittal Split Ramus Osteotomy Applied to Patients with Mandibular Asymmetry Using Three-Dimensional Finite Element Analysis

Cyprus J Med Sci 2025;10(1):44-50
1. Department of Oral and Maxillofacial Surgery Ankara University Faculty of Dentistry, Ankara, Türkiye
No information available.
No information available
Received Date: 23.05.2024
Accepted Date: 16.11.2024
Online Date: 14.03.2025
Publish Date: 14.03.2025
PDF
Cite
Share
Request

Abstract

BACKGROUND/AIMS

Fixation of the mobilized bone fragments is of importance for the healing of the patients and stabilization of the osteotomy in the postoperative period. In our study, models with different degrees of asymmetry were fixed with different fixation methods and the results were evaluated.

MATERIALS AND METHODS

Models with right rotations of 2, 5, and 10 mm were fixed with three bicortical screws and two miniplates with four monocortical screws, and the results were compared with those of finite element analysis (FEA). Tension and compression stresses in the bone segments and temporomandibular joint of two different fixation methods were compared.

RESULTS

FEA showed that the tensile and compressive stresses in the buccal, lingual, and temporomandibular discs were higher with bicortical screws.

CONCLUSION

It was predicted that stabilization problems would increase with increasing motion in sagittal split osteotomy. When selecting the most stable fixation method, the least stress to the surrounding tissues should be taken into consideration.

Keywords:
Three-dimensional finite element analysis, fixation systems, orthognathic surgery, sagittal split ramus osteotomy

INTRODUCTION

Mandibular osteotomies are surgical procedures performed on the ramus, corpus, and symphysis to treat mandible anomalies. Split osteotomy, which consists of osteotomies that separate the medial and lateral cortices of the ramus in the mandible.1 Sagittal split ramus osteotomy (SSRO) was first described by Trauner and Obwegeser2. Dalpont modified the technique in 1961 by positioning the buccal osteotomy behind the 2nd molars to increase the contact between the bone segments. In 1968, Hunsuck described the osteotomy procedure that is commonly used today, which involves terminating the lingual osteotomy just behind the lingula, extending to the posterior border of the ramus.1, 2

In SSRO, the mandible is segmented into two independent fragments, the proximal fragment and the distal fragment. In this technique, the distal bone fragment, which becomes free, is moved in a 3-dimensional plane. With a pre-prepared guide plate (occlusal splint) placed on the maxilla, the mandible is positioned in the required position and fixed with an appropriate fixation method.3, 4 Osteosynthesis methods in rigid internal fixation used in maxillofacial surgery consist of titanium mini-plates and monocortical and bicortical screws.5

The most important factors affecting the success of SSRO are stabilization, fixation instruments, and passive fusion of the bone fragments. As the movement speed increases, the changes in neighboring tissues and the stress on the tissues and fixation instruments also increase. Increased stress can also lead to changes in the sizes of miniplates, screws, and tissues.

Patients with mandibular asymmetry are treated with SSRO by changing the facial view, occlusion, and masticatory mechanics. With changes in the position of the mandible, the temporomandibular joint disk changes over time. The fixation methods used in osteotomy of the mandible lead to stress in the mandibular fragments and the temporomandibular joint. The stress on the mandible and temporomandibular joint increases with an increase in the amount of movement, which may lead to postoperative relapse, resorption, and temporomandibular joint disorders. The lack of a definite consensus on fixation methods directs surgeons to perform fixation based on observation and experience during the operation.6

Three-dimensional finite element analysis (FEA) and two-point biomechanical testing are the most important in vitro study designs developed to examine the reliability and effectiveness of fixation techniques. The FEA technique is a research method that can simulate the complex biomechanical analysis of the mandible close to reality and can change the direction and intensity of the forces applied on the model.7 In studies conducted with three-dimensional FEA, both time and cost savings are achieved, and the results obtained in simulations are reported to be compatible with clinical studies.8

In this study, models with different degrees of asymmetry will be fixed with different fixation methods using the Hunsuck method with SSRO and right rotation of the mandible model. The results will be evaluated by performing static linear analysis with the three-dimensional finite element method on models with 2 mm, 5 mm, and 10 mm right rotation in three different cases. This study aimed to compare the stress caused by different fixation methods on the bone segments and temporomandibular joint in different magnitudes of rotation movements compared with the FEA method.

MATERIALS AND METHODS

This study was conducted at the Ankara University. The Declaration of Helsinki was complied with, and approval was obtained from Ankara University Faculty of Dentistry Ethics Committee (approval number: 36290600/42/2023, date: 19.06.2023). This study was designed in the format of FEA, and the materials used in the study were provided through the Human Visible Project. No volunteer data were used, and no consent form was required.9

In this study, six three-dimensional models with SSRO were fixed with two bilateral, four-monocortical screws and flat miniplates and three bicortical screws after 2 mm, 5 mm and 10 mm rotation movements. The effects on the buccal and lingual segments and the temporomandibular joint were then examined using three-dimensional modeling and FEA.

In this study, the tensile and compression forces on various regions of the mandible caused by the plate and screw systems applied for fixation were evaluated. The analysis focuses solely on the forces and changes resulting from the plates and screws without any external forces being applied to the mandible.

The tomography images needed to create the three-dimensional models in the study were obtained from the Human Visible Project conducted by the US National Library of Medicine.9 The tomography images obtained from the Visible Human Project were scanned with a cross-sectional interval of 1 mm, and the “.stl” files of the three-dimensional toothed mandible model were transferred to 3D-Doctor (Able Software Corp., MA, USA) software in DICOM 3.0 format.9 After simplification and reformatting processes were applied to the images with 3D-Doctor software, the images were transferred to Rhinoceros 4.0 (3670 Woodland Park Ave N, Seattle, WA 98103 USA) 3D modeling software for 3D modeling. In the Rhinoceros program, osteotomies in accordance with the Hunsuck technique were simulated on the mandible model and 2 mm, 5 mm and 10 mm rotation movements were made to the distal segment. After creating the models of the plates and screws, the models were placed in the correct coordinates in three-dimensional space, and the modeling process was completed. The models created using Rhinoceros software were converted into geometric models, and a mesh structure was created using VR Mesh Studio (Virtual Grid Inc. Bellevue City, WA, USA) software to prepare them for analysis. Rhinoceros 4.0 (3670 Woodland Park Ave N, Seattle, WA 98103 USA) 3D modeling software, VRMesh Studio (VirtualGrid Inc, Bellevue city, WA, USA), and Algor Fempro (ALGOR, Inc. 150 Beta Drive Pittsburgh, PA 15238-2932 USA) analysis software was used for editing and homogenizing the 3D mesh structure, creating the 3D solid model, and finite element stress analysis.

The physical properties of the modeled cortical bone (Erkmen et al.14), spongiose bone (Erkmen et al.14), tooth (Ammoury et al.15), Ti-6 Al-4V (Bataineh and Janaideh16, Shu et al.18), zygomatic process (Mirow et al.17), articular cartilage (Mirow et al.17), temporomandibular ligament (Li et al.19), and disc (Li et al.19) are shown in Table 1.

Statistical Analysis

Evaluation of Finite Element Stress Analysis Results

Since the values obtained as a result of finite element stress analysis are the result of mathematical calculations without variance, statistical analysis cannot be performed. The precise evaluation and interpretation of the amount and distribution of stress in cross-sectional images and nodes is important.

In the analysis results, positive and negative values indicate tensile and compressive stresses, respectively. If the absolute value of a stress type is greater for a stress element, the stress element is under the influence of that stress type and should be evaluated.

The models with fixation with three linearly placed bicortical screws and two double-sided mini-plates with four monocortical screws and flat features are shown in Figure 1.

RESULTS

The maximum principal stress (tensile) values in the cortical bone are shown in the images on a scale ranging from red to blue, and the minimum principal stress (compression) values in the cortical bone are shown on a scale ranging from blue to red, as they are negative values.

The tensile and compressive stresses in the bone, which were rotated 10 mm to the right and fixed with two mini plates with bilateral, four-hole, and flat features, are shown in Figure 2, Graph 1, Table 2.

The tensile and compressive stresses in the bone, which were rotated 10 mm to the right and fixed with bicortical screws, are shown in Figure 3, Graph 2, Table 3.

DISCUSSION

SSRO is a popular osteotomy method for the correction of maxillofacial deformities and esthetic and functional incompatibilities of the mandible and maxilla. Fixation is important for the success of osteotomy to ensure the stabilization of bone fragments and healing. The ideal fixation should provide the highest stability between the bone fragments, provide adequate resistance to displacement forces at the osteotomy site, and place less stress on the surrounding tissues to ensure proper healing.6 As a result of the studies, the rigid internal fixation method has become the standard method, thus enabling patients to regain their postoperative functions more easily, and stabilization and relapse problems have been greatly reduced. In the rigid internal fixation class, postoperative complications are minimized using bicortical screws, monocortical screws, or combinations of the two, and fragment stabilization can be achieved.10, 11 Miloro13 demonstrated in their study that the use of three bicortical screws for the fixation of SSRO was stable, with an average relapse rate of 0-8%. Sigua-Rodriguez et al.11 reported that when two separate mini-plates are positioned in the tension and compression areas during fixation, the system is sufficiently resistant to displacement, resulting in lower stress at the osteotomy sites. In our study, we aimed to comparatively evaluate the fixation methods of three bicortical screws, which are considered more stable in the literature, and two mini-plates, which have been shown to generate less stress, using FEA.11-13Although rigid fixation with mini-plates and lag screws has several advantages, disadvantages such as nerve damage due to compression and displacement of the condyle from the fossa have also been reported.6 In order to avoid these disadvantages of rigid fixation, there are also studies that mention semi-rigid fixation.12 Mavili et al.12 investigated the short- and long-term stability of their semi-rigid fixation method in 23 patients who underwent maxillofacial surgery and 12 patients who underwent mandibular regression surgery. It was reported that the semi-rigid fixation method consisting of two bicortical screws with a diameter of 2 mm for mandibular fixation provided adequate stability and no recurrence in the short- and long-term. While fixation was achieved with bicortical screws, care was taken to maintain the gap between the fragments to avoid loading on the temporomandibular joint, and bone grafts were used when necessary. To stabilize semi-rigid fixation, jaw movements of the patients were restricted with maxillomandibular elastics for 2-4 weeks after 48 hours postoperatively. In our study, three bicortical screws and two miniplates, which are rigid fixation methods, were compared. The graft model was not placed in the opening between the fragments as the mandible moved. The fragments could approach each other under the forces applied by the screws. It was observed that as the amount of movement of the mandible increased, the gap between the fragments also increased. It was observed that the bicortical screws, while providing rigid fixation on the one hand, tended to close the gap between the buccal and lingual segments more with the compression force applied, thus creating more stress.

SSRO is a frequently preferred osteotomy method for the correction of maxillofacial deformities and esthetic and functional incompatibilities of the mandible and maxilla. Fixation is important for the success of osteotomy to ensure the stabilization of bone fragments and healing. The ideal fixation should provide the highest stability between the bone fragments, provide adequate resistance to displacement forces at the osteotomy site, and place less stress on the surrounding tissues to ensure proper healing.6 In our study, the stress and compression values of bicortical screws were higher than those of miniplate fixation models. In cases in which the bone fragments are flimsy and thin, it may be recommended to prefer plates because unwanted fractures may occur during surgery with the stress caused by bicortical screws. In the long term, resorption and temporomandibular joint problems are more likely to occur in patients who are fixed with bicortical screws than with miniplate. Sato et al.6 used five different rigid fixation techniques with miniplate and bicortical screws in a clinical setting. They fixed bone segments using rigid fixation methods in models that were advanced 5 mm with sagittal split osteotomy. They compared the stabilization values obtained by applying a force on the first molar tooth and evaluated these results using FEA. They reported that FEA is a numerical method for evaluating biomechanical problems and a powerful research tool that can provide precise information about the stress behavior of the mandible affected by mechanical forces. They reported that the mechanical connections between the distal and proximal segments, fixation materials, and stress in adjacent areas can be measured using this method. In the study, it was reported that the bicortical screw had higher stability than the mini plate. They explained that bicortical screws provide better three-dimensional stabilization than mini-plates because they attach to both bone segments. The mini-plates provide stabilization with a structure of bridge between the screws that provides force transmission, thus allowing flexion against external torsional forces. A force fracture transmits a lower compressive force to the bone fragments and lower torque force to the temporomandibular joint. Therefore, mini-plates have been reported as an alternative to bicortical screws in terms of biomechanical strength and transmitted stresses in sagittal split osteotomies. In the present study, we aimed to evaluate the stresses that may occur between segments using different fixation methods using FEA. It was found that the compression and tensile forces in the models fixed with the miniplate used were lower than bicortical screws. This may be explained by the fact that the force-transmitting plate allows flexion and force fracture. The stresses caused by the transmission of lower torque force to the temporomandibular joint were also lower in models with mini-plate fixation. Our results support the view that mini-plates are a good alternative to bicortical screws because of their lower stress generation.

FEA is a powerful in vitro method that can provide highly accurate information regarding the biomechanical behavior of the mandible, which exhibits diverse and complex properties. This method allows for the definition of different material characteristics and the modification of the magnitude and direction of the applied forces in the designed models. Three-dimensional FEA reflects the stress behavior on the models in a manner that is closer to reality than other in vitro methods, taking into account the complexities of clinical conditions.20The objective of studies conducted using the FEA method is to predict the behavior of designed biomaterials and existing systems, thereby guiding surgeons’ clinical decisions in a more predictable manner. In this study, we chose to employ this method for the aforementioned reasons.

Study Limitations

There are some limitations to this study, including the FEA. Although the mandibular bone model created by three-dimensional FEA is defined as isotropic, homogeneous, and linear elastic, the mandible is anisotropic and heterogeneous. Anatomically, each mandible has a different cortical bone density, spongiosis bone density, and masticatory-occlusive mechanics. The deformation characteristics and resorption patterns of each patient’s mandible differ according to external effects. A standardized and homogeneous mandible can be projected by FEA. In spite of these disadvantages, it is often preferred in scientific research due to its advantages, such as imitating biomechanically complex structures as close as possible to reality, changing the intensity and direction of the forces to be applied, and defining different material properties. The results obtained in this study need to be supported by clinical studies.

CONCLUSION

After sagittal split osteotomy, the mandibular models were rotated 2 mm, 5 mm, and 10 mm to the right and fixed using two different methods: three bicortical screws in a linear position and two mini plates with four screws. After fixation, the tension and compression stresses on the buccal and lingual bone segments on the right and left sides of the mandible and condyle were analyzed. The stress on the condyle and bone fragments increased as the movement speed increased. When we compared the two different fixation techniques, the stresses on the bone segments and condyle were higher in fixation with bicortical screws.

It was predicted that stabilization issues would increase with increasing the amount of motion during sagittal split osteotomy. When selecting the most stable fixation method, the least stress to the surrounding tissues should be taken into consideration.

MAIN POINTS

• Stabilization of the fixation is an important factor affecting the success of sagittal split osteotomy.

• Bicortical screws, monocortical screws, and plates are commonly used for rigid internal fixation.

• Bicortical screws used during fixation may cause higher tensile and compressive stresses than monocortical screws and plates.

ETHICS

Ethics Committee Approval: The Declaration of Helsinki was complied with, and approval was obtained from Ankara University Faculty of Dentistry Ethics Committee (approval number: 36290600/42/2023, date: 19.06.2023).

Informed Consent: Not applicable.

Authorship Contributions

Surgical and Medical Practices: M.Ö., Concept: M.Ö., S.A., Design: M.Ö., S.A., Data Collection and/or Processing: M.Ö., S.Ö.K., Analysis and/or Interpretation: M.Ö., Literature Search: M.Ö., S.Ö.K., Writing: M.Ö., S.Ö.K.

Conflict of Interest: No conflict of interest was declared by the authors.

Financial Disclosure: The authors declared that this study had received no financial support.

Ethics

Ethics Committee Approval: The Declaration of Helsinki was complied with, and approval was obtained from Ankara University Faculty of Dentistry Ethics Committee (approval number: 36290600/42/2023, date: 19.06.2023).
Informed Consent: Not applicable.

Authorship Contributions

Surgical and Medical Practices: M.Ö., Concept: M.Ö., S.A., Design: M.Ö., S.A., Data Collection and/or Processing: M.Ö., S.Ö.K., Analysis and/or Interpretation: M.Ö., Literature Search: M.Ö., S.Ö.K., Writing: M.Ö., S.Ö.K.
Conflict of Interest: No conflict of interest was declared by the authors.
Financial Disclosure: The authors declared that this study had received no financial support.

References

1
Cevidanes LH, Hajati AK, Paniagua B, Lim PF, Walker DG, Palconet G, et al. Quantification of condylar resorption in temporomandibular joint osteoarthritis. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2010; 110(1): 110-7.
2
Trauner R, Obwegeser H. The surgical correction of mandibular prognathism and retrognathia with consideration of genioplasty. I. Surgical procedures to correct mandibular prognathism and reshaping of the chin. Oral Surg Oral Med Oral Pathol. 1957; 10(7): 677-89.
3
Ghang MH, Kim HM, You JY, Kim BH, Choi JP, Kim SH, et al. Three-dimensional mandibular change after sagittal split ramus osteotomy with a semirigid sliding plate system for fixation of a mandibular setback surgery. Oral Surg Oral Med Oral Pathol Oral Radiol. 2013; 115(2): 157-66.
4
Gursoytrak B, Unsal N, Demetoglu U, Simsek HO, Saglam H, Dolanmaz D. Biomechanical evaluation of hybrid fixation method of sagittal split ramus osteotomy in mandibular advancement. J Craniomaxillofac Surg. 2018; 46(12): 2063-8.
5
Brasileiro BF, Grempel RG, Ambrosano GM, Passeri LA. An in vitro evaluation of rigid internal fixation techniques for sagittal split ramus osteotomies: advancement surgery. J Oral Maxillofac Surg. 2009; 67(4): 809-17.
6
Sato FR, Asprino L, Noritomi PY, da Silva JV, de Moraes M. Comparison of five different fixation techniques of sagittal split ramus osteotomy using three-dimensional finite elements analysis. Int J Oral Maxillofac Surg. 2012; 41(8): 934-41.
7
Küçükkurt S. Finite element stress analysis method and research related to dental implantology. Journal of Atatürk University Faculty of Dentistry. 2018; 29(4): 701-10.
8
Maurer P, Knoll WD, Schubert J. Comparative evaluation of two osteosynthesis methods on stability following sagittal split ramus osteotomy. J Craniomaxillofac Surg. 2003; 31(5): 284-9.
9
Ackerman MJ. The Visible Human Project: a resource for education. Acad Med. 1999; 74(6): 667-70.
10
Peterson GP, Haug RH, Van Sickels J. A biomechanical evaluation of bilateral sagittal ramus osteotomy fixation techniques. J Oral Maxillofac Surg. 2005; 63(9): 1317-24.
11
Sigua-Rodriguez EA, Caldas RA, Goulart DR, Hemerson de Moraes P, Olate S, Ricardo Barão VA, et al. Comparative evaluation of different fixation techniques for sagittal split ramus osteotomy in 10 mm advancements. Part two: finite element analysis. J Craniomaxillofac Surg. 2019; 47(7): 1015-9.
12
Mavili ME, Canter HI, Saglam-Aydinatay B. Semirigid fixation of mandible and maxilla in orthognathic surgery: stability and advantages. Ann Plast Surg. 2009; 63(4): 396-403.
13
Miloro M. Principles of mandibular orthognathic surgery. In Principles of Oral and Maxillofacial Surgery. 2004; 1135-78.
14
Erkmen E, Simşek B, Yücel E, Kurt A. Comparison of different fixation methods following sagittal split ramus osteotomies using three-dimensional finite elements analysis. Part 1: advancement surgery-posterior loading. Int J Oral Maxillofac Surg. 2005; 34(5): 551-8.
15
Ammoury MJ, Mustapha S, Dechow PC, Ghafari JG. Two distalization methods compared in a novel patient-specific finite element analysis. Am J Orthod Dentofacial Orthop. 2019; 156(3): 326-36.
16
Bataineh K, Al Janaideh M. Effect of different biocompatible implant materials on the mechanical stability of dental implants under excessive oblique load. Clin Implant Dent Relat Res. 2019; 21(6): 1206-17.
17
Mirow E, Sifakakis I, Keilig L, Bourauel C, Patcas R, Eliades T, et al. Quantitative appraisal of bilateral sagittal split osteotomy impact on the loading of temporomandibular joint. J Mech Behav Biomed Mater. 2020; 111: 103985.
18
Shu J, Zhang Y, Liu Z. Biomechanical comparison of temporomandibular joints after orthognathic surgery before and after design optimization. Med Eng Phys. 2019; 68: 11-6.
19
Li H, Zhou N, Huang X, Zhang T, He S, Guo P. Biomechanical effect of asymmetric mandibular prognathism treated with BSSRO and USSRO on temporomandibular joints: a three-dimensional finite element analysis. Br J Oral Maxillofac Surg. 2020; 58(9): 1103-9.
20
Türker N, Büyükkaplan US, Sadowsky SJ, Özarslan MM. Finite element stress analysis of applied forces to implants and supporting tissues using the “all-on-four” concept with different occlusal schemes. J Prosthodont. 2019; 28(2): 185-94.